Title: Sequential determination of iron and titanium by flow-injection analysis
Other Titles: Talanta
Authors: Barberino, Maria Goreth Matos de Andrade
Ferreira, Sergio Luis Costa
Santos, B. F.
Costa, Antonio Celso Spinola
metadata.dc.creator: Barberino, Maria Goreth Matos de Andrade
Ferreira, Sergio Luis Costa
Santos, B. F.
Costa, Antonio Celso Spinola
Abstract: flow-injection method has been developed for the sequential spectrophotometric determination of iron and titanium using 3,s dihydroxybenzoic acid as chromogenic reagent. The system involves the sequential measurement of the absorbances of the complexes at 380 and 570 nm. The system is designed using a simultaneous injection of sample and reagent into separate carrier streams. The proposed method is characterized by a precision of about 2%, a sampling rate of about 50 samples per hour, and a reagent consumption of 200 gl(O.50% solution) per sample. It is relatively free of interferences and was used for the sequential determination of titanium and iron in rocks.
Publisher: Talanta
URI: http://www.repositorio.ufba.br/ri/handle/ri/7395
Issue Date: 1992
Appears in Collections:Artigo Publicado em Periódico (Química)