Please use this identifier to cite or link to this item: https://repositorio.ufba.br/handle/ri/4891
metadata.dc.type: Artigo de Periódico
Title: Um espectômetro de massas de tempo-de-vôo para o estudo da ionização de amostras gasosas por elétrons rápidos (0,5 – 1,0 KEV)
Other Titles: Química Nova
Authors: Cardoso, Aldo S.
Pontes, Frederico C.
Souza, Gerardo Gerson Bezerra de
Mundim, Maria Suely Pedrosa
metadata.dc.creator: Cardoso, Aldo S.
Pontes, Frederico C.
Souza, Gerardo Gerson Bezerra de
Mundim, Maria Suely Pedrosa
Abstract: A TIME-OF-FLIGHT MASS SPECTROMETER FOR HIGH ENERGY (0.5 – 1.0 KEV) ELECTRON IONIZATION OF GASES. A simple and inexpensive time-of-flight mass spectrometer, dedicated to the study of gas-phase ionization processes induced by high energy electrons (0.5 – 3.0 keV), is described. The spectrometer design is based on the Wiley-McLaren principle, with a total length of about 18 cm. As a demonstration of the performance of the apparatus, mass spectra for Ar, CH4, CO2, and SF6, obtained at 1 keV electron energy, are presented.
Keywords: Time-of-flight mass spectrometer
Electron ionization
multiply-charged ions of argon
URI: http://www.repositorio.ufba.br/ri/handle/ri/4891
Issue Date: 2001
Appears in Collections:Artigo Publicado em Periódico (FIS)

Files in This Item:
File Description SizeFormat 
a06v24n3.pdf71,86 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.