Use este identificador para citar ou linkar para este item: https://repositorio.ufba.br/handle/ri/13144
Tipo: Artigo de Periódico
Título: A photo-oxidation procedure using UV radiation/H2O2 for decomposition of wine samples: determination of iron and manganese content by flame atomic absorption spectrometry
Título(s) alternativo(s): Spectrochimica Acta Part B: Atomic Spectroscopy
Autor(es): Santos, Walter Nei Lopes dos
Brandão, Geovani Cardoso
Portugal, Lindomar Andrade
David, Jorge Mauricio
Ferreira, Sergio Luis Costa
Autor(es): Santos, Walter Nei Lopes dos
Brandão, Geovani Cardoso
Portugal, Lindomar Andrade
David, Jorge Mauricio
Ferreira, Sergio Luis Costa
Abstract: This paper proposes the use of photo-oxidation with UV radiation/H2O2 as sample pretreatment for the determination of iron and manganese in wines by flame atomic absorption spectrometry (FAAS). The optimization involved the study of the following variables: pH and concentration of buffer solution, concentrated hydrogen peroxide volume and irradiation time. The evaluation of sample degradation was monitored by measuring the absorbance at the maximum wavelength of red wine (530 nm). Using the experimental conditions established during the optimization (irradiation time of 30 min, oxidant volume of 2.5 mL, pH 10, and a buffer concentration of 0.15 mol L− 1), this procedure allows the determination of iron and manganese with limits of detection of 30 and 22 μg L− 1, respectively, for a 5 mL volume of digested sample. The precision levels, expressed as relative standard deviation (RSD), were 2.8% and 0.65% for iron and 2.7% and 0.54% for manganese for concentrations of 0.5 and 2.0 mg L− 1, respectively. Addition/recovery tests for evaluation of the accuracy were in the ranges of 90%–111% and 95%–107% for iron and manganese, respectively. This digestion procedure has been applied for the determination of iron and manganese in six wine samples. The concentrations varied from 1.58 to 2.77 mg L− 1 for iron and from 1.30 to 1.91 mg L− 1 for manganese. The results were compared with those obtained by an acid digestion procedure and determination of the elements by FAAS. There was no significant difference between the results obtained by the two methods based on a paired t-test (at 95% confidence level).
Palavras-chave: Photo-oxidation procedure
UV radiation/H2O2
Sample decomposition
Wine samples
Metal determination
FAAS
URI: http://www.repositorio.ufba.br/ri/handle/ri/13144
Data do documento: 2009
Aparece nas coleções:Artigo Publicado em Periódico (Química)

Arquivos associados a este item:
Não existem arquivos associados a este item.


Os itens no repositório estão protegidos por copyright, com todos os direitos reservados, salvo quando é indicado o contrário.